FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4031502 · Received August 22, 2014

Report

Report Number
1416980-2014-27732
Event Type
Malfunction
Date Received
August 22, 2014
Date of Event
July 31, 2014
Report Date
July 31, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN AS PART OF PREVENTIVE MAINTENANCE. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, POWER ON SELF-TESTED AND THE ALARM LOG AND SERVICE HISTORY LOG WERE REVIEWED. DURING POWER ON SELF-TEST AND ALARM LOG REVIEW THE F-38 ALARM WAS FOUND; THE CAUSE OF THIS WAS DAMAGED FORCE SENSING RESISTORS (FSR). TO CORRECT THE CONDITION, THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A FIELD SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509208 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1