UV FLASH TRANSFER SETS
Report
- Report Number
- 1416980-2014-27730
- Event Type
- Malfunction
- Date Received
- August 22, 2014
- Report Date
- July 30, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE EVALUATION IS COMPLETE. THE DAMAGED PATIENT CONNECTOR WAS IDENTIFIED DURING VISUAL INSPECTION. LEAK (UNDER WATER, USING PRESSURE), CLEAR PASSAGE, AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED; THE DEVICE PASSED THESE TESTS. THE CAUSE WAS UNABLE TO BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED. THIS IS THE SAME DEVICE AS (B)(4).
(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE INVESTIGATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED UV FLASH TRANSFER SET, THE PATIENT CONNECTOR WAS NOTED TO BE DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509156 | UV FLASH TRANSFER SETS | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |