FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 4030548 · Received August 22, 2014

Report

Report Number
2955842-2014-05135
Event Type
Malfunction
Date Received
August 22, 2014
Date of Event
July 10, 2014
Report Date
July 24, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE. THE INSTRUMENT INSTALLED ON AN IN-HOUSE IS3000 SYSTEM PASSED INTUITIVE MOTION TESTING. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE DISTAL END OF THE INSTRUMENT'S MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES ARE .021 - .231 IN LENGTH AND ARE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE ADDITIONAL FINDING OF TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HEMICOLECTOMY PROCEDURE, AFTER THE PROGRASP FORCEPS INSTRUMENT WAS INSTALLED ON THE SYSTEM, THE JAWS ON THE INSTRUMENT FROZE. THERE WAS NO REPORT THAT ANY PIECE(S) FROM THE INSTRUMENT FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
510266 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M10131005 653

Patients

Seq Age Sex Outcome Treatment
1