HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-27614
- Event Type
- Malfunction
- Date Received
- August 22, 2014
- Date of Event
- July 29, 2014
- Report Date
- July 29, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USER NOT TO ATTEMPT TO REUSE ANY DISPOSABLE SUPPLIES. IT WARNS THE USER NOT TO REPLACE EMPTY SOLUTION BAGS OR RECONNECT DISCONNECTED SOLUTION BAGS DURING THERAPY. ALSO, IT WARNS THE USER THAT POSSIBLE CONTAMINATION OF THE FLUID OR FLUID PATHWAYS CAN RESULT IF DISPOSABLES ARE REUSED. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING TROUBLESHOOTING OF AN UNRELATED ALARM, IT WAS NOTED THAT THE EMPTY HEATER BAG WAS DISCONNECTED AND A NEW SOLUTION BAG WAS CONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER PROCEDURES WITH THE CARE GIVER (CG) AND ASSISTED IN ENDING THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509138 | HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR | HOMECHOICE |