FDA Adverse Event Malfunction Summary report: N

STATSPIN MP

MDR report key: 4030377 · Received July 29, 2014

Report

Report Number
2023446-2014-00096
Event Type
Malfunction
Date Received
July 29, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
IRIS DIAGNOSTICS
Product Code
GKG
PMA / PMN Number
K920759
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER REPORTED A BURNING SMELL AND VISIBLE SMOKE WHILE CENTRIFUGE WAS IN USE. UPON OPENING THE UNIT, IT WAS OBSERVED THAT THE O-RING HAD POPPED OUT OF THE ROTOR AND HAD MELTED INSIDE THE UNIT. THERE WAS NO HARM DUE TO THIS INCIDENT AND NO ONE REQUIRED MEDICAL ATTENTION. THERE WERE NO REPORTS OF VISIBLE FLAMES NOR FIRE DEPARTMENT BEING CALLED.

Description of Event or Problem · 1

CUSTOMER NOTED SMOKE COMING OUT OF THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
442553 STATSPIN MP GKG IRIS DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1