FDA Adverse Event
Malfunction
Summary report: N
CARE ASSIST BED
MDR report key: 4030366
·
Received July 18, 2014
Report
- Report Number
- 9615739-2014-00288
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 26, 2014
- Report Date
- June 26, 2014
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECH FOUND FOUR BRAKE CASTERS AND CASTER SUPPORTS INOPERABLE MOST LIKELY DUE TO WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTIVE MAINTENANCE ON THIS BED IN 2012 AND 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH REPLACED FOUR BRAKE CASTERS AND CASTER SUPPORTS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THAT THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN MAINTENANCE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).