FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 4025115 · Received August 20, 2014

Report

Report Number
2245578-2014-00043
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 25, 2014
Report Date
July 30, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 09/08/2014. RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.49 ON A PATIENT WHO PRESENTED WITH CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.DATE SAMPLE DRAW TIME RESULT TIME METHOD RESULT (NG/ML) SAMPLE(B)(6) 2014 10:23 10:28 I-STAT 0.49 A(B)(6) 2014 10:23 10:43 I-STAT 0.02 A(B)(6) 2014 10:42 10:52 I-STAT 0.04 BTHERE WERE NO INJURIES REPORTED WITH THIS EVENT. AN INVESTIGATION HAS BEEN INITIATED.