I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2014-00043
- Event Type
- Malfunction
- Date Received
- August 20, 2014
- Date of Event
- July 25, 2014
- Report Date
- July 30, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE INVESTIGATION WAS COMPLETED ON 09/08/2014. RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
(B)(4).
ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.49 ON A PATIENT WHO PRESENTED WITH CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.DATE SAMPLE DRAW TIME RESULT TIME METHOD RESULT (NG/ML) SAMPLE(B)(6) 2014 10:23 10:28 I-STAT 0.49 A(B)(6) 2014 10:23 10:43 I-STAT 0.02 A(B)(6) 2014 10:42 10:52 I-STAT 0.04 BTHERE WERE NO INJURIES REPORTED WITH THIS EVENT. AN INVESTIGATION HAS BEEN INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 503356 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE | NA | P14136 |