MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 4025017 · Received August 20, 2014

Report

Report Number
2955842-2014-05083
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 9, 2014
Report Date
July 25, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE SCISSORS DID NOT FULLY CLOSE DUE TO BLADE DAMAGE. AS A RESULT, THE BLADES WOULD NOT CUT THROUGH THEIR ENTIRE LENGTH AND A CUT TEST FAILED. ONE BLADE EDGE HAD AN INDENTATION BETWEEN THE MIDPOINT AND TIP THAT ACTED AS A MECHANICAL STOP FOR THE OTHER BLADE WHEN CLOSING. IT WAS CONCLUDED THAT THE INDENTATION DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE APPROXIMATELY .032 - .078 IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE SCISSORS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT WERE BLUNT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
503591 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-12 N11140424 637

Patients

Seq Age Sex Outcome Treatment
1 64 YR