FDA Adverse Event Malfunction Summary report: N

TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM

MDR report key: 4024931 · Received August 20, 2014

Report

Report Number
2183870-2014-00212
Event Type
Malfunction
Date Received
August 20, 2014
Date of Event
July 22, 2014
Report Date
July 22, 2014
Manufacturer
COVIDIEN
Product Code
MCW
PMA / PMN Number
K103618
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURE RECORDS FOR THIS DEVICE DID NOT REVEAL ANY DISCREPANCIES RELEVANT TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN PLYMOUTH LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.

Description of Event or Problem · 1

FIVE PASSES WERE MADE IN THE COMMON FEMORAL. AS THE TURBOHAWK WAS BEING REMOVED THE NOSE CONE CAME OFF THE CATHETER AND WAS STUCK ON THE WIRE INSIDE THE SHEATH. THE PHYSICIAN WAS ABLE TO REMOVE THE WIRE WITH THE NOSE CONE OUT OF THE BODY AND THE TECH HELD MANUAL PRESSURE.

Description of Event or Problem · 1

DEVICE EVALUATION: THE DISTAL TIP ASSEMBLY WAS FRACTURED AT THE PROXIMAL EDGE OF THE HOUSING (LASER-CUT) COIL. THE PROXIMAL HALF OF THE GUIDEWIRE LUMEN (GWL) WAS PULLED THROUGH THE TECOTHANE SLEEVE AND WAS TORN LONGITUDINALLY. THE REMAINING GUIDEWIRE LUMEN EXHIBITED A LONGITUDINAL TEAR BUT REMAINED ENCASED IN THE TECOTHANE SLEEVE UP THE ROTATING TIP. THE PROXIMAL 2MM OF THE ROTATING TIP'S GWL EXHIBITED SIMILAR DAMAGE. THE CUTTER HEAD ASSEMBLY EXHIBITED A FILM OF CLEAR POLYMER CONSISTENT IN APPEARANCE AND TEXTURE TO THE PET LINER OF THE DISTAL TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
501358 TURBOHAWK PERIPHERAL PLAQUE EXCISION SYSTEM CATHETER, PERIPHERAL, ATHERECTOMY MCW COVIDIEN THS-LS-C 9838516

Patients

Seq Age Sex Outcome Treatment
1 69 YR Other 7F CROSSOVER SHEATH