FDA Adverse Event Injury Summary report: N

VERION DIGITAL MARKER M FOR ZEISS MICROSCOPES

MDR report key: 4024527 · Received August 1, 2014

Report

Report Number
3010300699-2014-00002
Event Type
Injury
Date Received
August 1, 2014
Date of Event
June 4, 2014
Report Date
July 4, 2014
Manufacturer
WAVELIGHT GMBH (AGPS)
Product Code
FTH
PMA / PMN Number
EXERR
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. ADDITIONAL INFO HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE. (B)(4).

Description of Event or Problem · 1

AN OPHTHALMIC SURGEON REPORTED THAT AFTER CATARACT SURGERY SHE IDENTIFIED A DEVIATION IN THE AXIS FOR THE IMPLANTED TORIC IOL IN THREE PATIENTS. DURING SURGERY, SHE SAW A ROTATION OF THE FINAL AXIS COMPARED WITH THE INITIALLY PLANNED AXIS. THIS LED TO RESIDUAL ASTIGMATISM THAT WAS CORRECTED WITH EYE GLASSES. FOR THIS PT SURGEON REPORTED A JUMP ON THE MARKS FOR THE INCISION OF 30 DEGREES. THE SURGEON INDICATED THE POSTOPERATIVE ASTIGMATISM WAS DUE TO THE SYSTEM MAKINGS AND NOT DUE TO A POSTOPERATIVE IOL ROTATION. ADDITIONAL INFO WAS RECEIVED FROM THE SURGEON. DURING SURGERY, MARKINGS IN THE MICROSCOPE AND IN THE DISPLAY WERE PERMANENTLY CHANGING, THUS MAKING IT DIFFICULT TO GUESS WHICH WAS THE CORRECT TORIC ALIGNMENT AXIS. THE SURGEON CHOSE ONE OF THE MULTIPLE AXIS DISPLAYED AND POSTOPERATIVELY THE IOL PRESENTED A ROTATION OF 12 DEGREES DEVIATION FROM THE PREOPERATORY AXIS SUGGESTED BY THE DIGITAL MARKER (164 DEGREE INSTEAD OF 176 DEGREE). THE ROTATION WAS OBSERVED ON THE FIRST POSTOPERATIVE DAY AND PERSISTED ONE MONTH AFTER SURGERY. THE PT WAS NOT WILLING TO UNDERGO IOL REPOSITIONING. ADDITIONAL INFO HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451645 VERION DIGITAL MARKER M FOR ZEISS MICROSCOPES CAMERA, STILL, MICROSURGICAL FTH WAVELIGHT GMBH (AGPS) X-SPM NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention AVODART| UNK ACRYSOF TORIC IOL| IZODIENT| ATENOLOL