ENSITE VELOCITY SYSTEM
Report
- Report Number
- 2184149-2014-00005
- Event Type
- Death
- Date Received
- August 1, 2014
- Date of Event
- July 9, 2014
- Report Date
- July 9, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC (AF-ST. PAUL)
- Product Code
- DQK
- PMA / PMN Number
- K130594
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
TWO DISKS WERE RETURNED CONTAINING THE CASE STUDY AND LOGS. NO RECORDED SEGMENTS WERE MADE OF THE REPORTED EVENT ON EITHER OF THE RETURNED DISK. SYNCHRONIZED CARDIOVERSION CAN CAUSE A TRANSIENT CATHETER SHIFT. AFTER THE IMMEDIATE CIRCUMSTANCE HAS STABILIZED, THERE CAN BE A RESIDUAL CHANGE IN THORACIC IMPEDANCE THAT CAN HAVE AN IMPACT ON LOCATION. THIS EXPECTED BEHAVIOR OF THE VELOCITY SYSTEM IS RELATED TO CHANGES IN THE IMPEDANCE SIGNALS SENSED BY THE AMPLIFIER.
RELATED MANUFACTURER REFERENCE 3005188751-2014-00088, 3005188751-2014-00084, 3005188751-2014-00089, 2030404-2014-00077, 2030404-2014-00078, 2030404-2014-00079. FOLLOWING A VENTRICULAR TACHYCARDIA (VT) ABLATION PROCEDURE, THE PATIENT EXPIRED. THIS HIGH RISK SUBSTRATE MODIFICATION VT ABLATION PROCEDURE IN THE LEFT VENTRICLE WAS PERFORMED ON AN ELDERLY MALE WITH SEVERE CARDIOMYOPATHY, A HISTORY OF LARGE LAD TERRITORY MYOCARDIAL INFARCTION, AND AN IMPLANTED CARDIAC PACEMAKER/DEFIBRILLATOR. THE PATIENT PRESENTED TO THE LAB IN A PACED RHYTHM AND ANESTHESIA WAS ADMINISTERED; THE PATIENT SUBSEQUENTLY BECAME HYPOTENSIVE. AN EPINEPHRINE INFUSION WAS INITIATED TO STABILIZE AND MAINTAIN BLOOD PRESSURE. TRANSSEPTAL PUNCTURE WAS PERFORMED WITH TRANSESOPHAGEAL ECHOCARDIOGRAM GUIDANCE USING A BRK TRANSSEPTAL NEEDLE AND AN AGILIS NXT INTRODUCER WAS ADVANCED INTO THE LEFT ATRIUM. FOLLOWING THE TRANSSEPTAL PUNCTURE, A PERICARDIAL EFFUSION WAS NOTED; HOWEVER, THIS WAS NOT SIGNIFICANT AND THE PROCEDURE CONTINUED. A NON-SJM SCREW IN BIPOLAR PACING CATHETER WAS PLACED IN THE RIGHT VENTRICLE AND AN INQUIRY AFOCUS II EP CATHETER WAS ADVANCED INTO THE LEFT VENTRICLE. THE AFOCUS II EP CATHETER WAS USED TO MAP, BOTH TRANSSEPTAL AND VIA A RETROGRADE APPROACH. THE AFOCUS II EP CATHETER WAS DAMAGED AND THEN REPLACED TO CONTINUE MAPPING. DURING COMPLETION OF THE MAP, THE PATIENT DEVELOPED VT, RESULTING IN HEMODYNAMIC INSTABILITY; THEREFORE, THE PATIENT WAS CARDIOVERTED TO SINUS RHYTHM (SR). THE CATHETERS SHIFTED ON THE ENSITE VELOCITY MAPPING SYSTEM DUE TO THE CARDIOVERSION AND SETTINGS CHANGES. RE-MAPPING WAS COMPLETED AND THE AFOCUS II EP CATHETER WAS REPLACED WITH A TACTICATH QUARTZ ABLATION CATHETER. CATHETER DISPLAY ISSUES WERE NOTED AND TROUBLESHOOTING WAS PERFORMED, EVENTUALLY RESOLVING THIS ISSUE. RE-MAPPING WAS PERFORMED AGAIN USING A THERAPY COOL FLEX ABLATION CATHETER. ABLATION WAS SUCCESSFULLY PERFORMED AROUND A VERY LARGE SCAR WHICH STRETCHED FROM THE APEX TO THE BASE OF THE LEFT VENTRICLE. NEAR THE CONCLUSION OF THE PROCEDURE, THE ANESTHESIA TEAM NOTED IT WAS DIFFICULT TO MAINTAIN A STABLE BLOOD PRESSURE AND THE PERICARDIAL EFFUSION NOTED EARLIER HAD INCREASED IN SIZE. PROTAMINE WAS ADMINISTERED AND A PERICARDIAL DRAIN WAS PLACED, WHICH STABILIZED THE PATIENT'S BLOOD PRESSURE. THE PATIENT WAS AWAKENED AND TRANSFERRED TO THE RECOVERY ROOM IN STABLE CONDITION. THE PATIENT THEN BECAME HYPOTENSIVE AND AN EPINEPHRINE BOLUS WAS ADMINISTERED. AN ECHOCARDIOGRAM REVEALED AN EFFUSION WITH NO LEFT VENTRICULAR MOVEMENT. THE PATIENT SUBSEQUENTLY EXPIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 451625 | ENSITE VELOCITY SYSTEM | COMPUTER, DIAGNOSTIC, PROGRAMMABLE | DQK | ST. JUDE MEDICAL, INC (AF-ST. PAUL) | VELOCITY |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Death| R | PN-004075/(B)(4)| MODEL UNK, LOT UNK| CARDIAC ABLATION GENERATOR CONNECTING CABLE:| 407356/(B)(4)| SWARTZ BRAIDED TRANSSEPTAL INTRODUCER:| INQUIRY AFOCUS II EP CATHETER: 87008/(B)(4)| 85809, LOT UNK| INQUIRY AFOCUS II EP CATHETER: 87008/(B)(4)| NON-SJM SCREW IN BIPOLAR PACING CATHETER:| AGILIS NXT INTRODUCER: G408332/(B)(4)| A088017/4422488| THERAPY COOL FLEX ABLATION CATHETER:| PN-004075/(B)(4)| TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER:| BRK TRANSSEPTAL NEEDLE: 407200/(B)(4)| TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER: |