FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 4021879 · Received August 19, 2014

Report

Report Number
2938836-2014-14453
Event Type
Injury
Date Received
August 19, 2014
Date of Event
July 12, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WAS RETURNED IN TWO SEGMENTS FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW-UP, NON-SUSTAINED VENTRICULAR OVERSENSING WAS OBSERVED. NO PATIENT SYMPTOMS WERE REPORTED. REVIEW OF THE EGMS REVEALED POSSIBLE LEAD NOISE. NOISE COULD NOT BE REPLICATED WITH PROVOCATIVE TESTING. LEAD IMAGING WAS RECOMMENDED.

Description of Event or Problem · 1

NEW INFORMATION RECEIVED NOTES THAT THE LEAD WAS EXPLANTED AND REPLACED DUE TO NOISE. THE PATIENT DID WELL DURING AND AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
497811 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7121/65 NA

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention