FDA Adverse Event Injury Summary report: N

COMP RVS HMRL TI TRAY 44MM

MDR report key: 4020255 · Received August 19, 2014

Report

Report Number
0001825034-2014-07150
Event Type
Injury
Date Received
August 19, 2014
Date of Event
July 30, 2014
Report Date
May 12, 2015
Manufacturer
BIOMET ORTHOPEDICS
Product Code
PAO
PMA / PMN Number
PK080642
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION THAT WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 10 STATES, "FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT."

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT A COMPREHENSIVE REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2009. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2014 DUE TO A FRACTURED HUMERAL TRAY AFTER HEAVY LIFTING. THE HUMERAL TRAY AND HUMERAL BEARING WERE REMOVED AND REPLACED. ADDITIONAL INFORMATION RECEIVED IN OPERATIVE REPORT NOTED PATIENT WAS REVISED ON (B)(6) 2014 DUE TO PAIN AND A FRACTURED HUMERAL BASEPLATE. THE HUMERAL COMPONENTS AND TAPER WERE REMOVED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT A COMPREHENSIVE REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2009. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2014 DUE TO A FRACTURED HUMERAL TRAY AFTER HEAVY LIFTING. THE HUMERAL TRAY AND HUMERAL BEARING WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498167 COMP RVS HMRL TI TRAY 44MM PROSTHESIS, SHOULDER PAO BIOMET ORTHOPEDICS N/A 607420

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| R