FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4020244 · Received August 19, 2014

Report

Report Number
2531779-2014-23826
Event Type
Malfunction
Date Received
August 19, 2014
Report Date
August 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/03/2014 WITH THE FOLLOWING FINDINGS: THE DAILY INSULIN DELIVERY TOTALS WERE REVIEWED AND FOUND TO CORRECTLY REFLECT THE PROGRAMMED BASAL RATES. A DELIVERY ACCURACY TEST WAS PERFORMED AND PASSED. THE PUMP PERFORMED WITHIN SPECIFICATIONS. NO DEFECT WAS FOUND.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (INACCURATE DELIVERY) ISSUE. IT WAS REPORTED THAT THE USER BLOOD GLUCOSE (BG) READING WAS GREATER THAN 250 MG/DL; HOWEVER, IT WAS NOT ALLEGED THAT THE BG READING EXCEEDED 500 MG/DL. NO SYMPTOMS OF HYPERGLYCEMIA WERE REPORTED. THE ALLEGED BG EXCURSION DOES NOT MEET ANIMAS CRITERIA FOR A SERIOUS INJURY AND IS THEREFORE NOT REPORTABLE AS AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
497958 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR