ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-23819
- Event Type
- Malfunction
- Date Received
- August 19, 2014
- Report Date
- August 10, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/12/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. ON INVESTIGATION, MOISTURE INTRUSION WAS EVIDENT INSIDE THE BATTERY COMPARTMENT. THE PUMP FAILED TO POWER UP WITH THE RETURNED OR TEST BATTERY CAP. BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED BELOW THE GRIP PAD. DUE TO THE NO POWER ISSUE, THE REMAINING TESTING CANNOT BE PERFORMED AND THE ALLEGED BATTERY LIFE ISSUE WAS NOT FULLY INVESTIGATED. NO CONCLUSION CAN BE DRAWN.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE BATTERY LIFE WAS LESS THAN EXPECTED. REVIEW OF ALARM HISTORY SHOWED EVIDENCE OF SHORT BATTERY LIFE. REPORTEDLY, THE ISSUE RECURRED WITH MULTIPLE BATTERIES FROM DIFFERENT PACKS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 499378 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |