FDA Adverse Event Malfunction Summary report: N

PRISMALIX

MDR report key: 4019925 · Received July 11, 2014

Report

Report Number
9710055-2014-00054
Event Type
Malfunction
Date Received
July 11, 2014
Report Date
June 11, 2014
Manufacturer
MAQUET S.A.S.
Product Code
FSY
PMA / PMN Number
K040015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

HOSPITAL STAFF SENT PHOTOS OF THE CHIPPED PAINT AS WELL AS THE CLEANING PROCEDURE USED, BUT THE PRECISE DEVICE SERIAL NUMBER WAS NOT PROVIDED. ON OF THE CLEANING AGENTS DESCRIBED IN THE HOSPITAL PROCEDURE CONTAINS PHENOL, WHICH IS IDENTIFIED IN THE DEVICE'S USER MANUAL AS A PROHIBITED CLEANING SUBSTANCE. MAQUET MEDICAL SYSTEMS USA SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MFG FACILITY. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THEIR SURGICAL LIGHT WAS STARTING TO EXPERIENCE FLAKING PAINT CHIPS. NO PATIENT INVOLVEMENT WAS REPORTED BY THE CUSTOMER. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
407314 PRISMALIX FSY MAQUET S.A.S. PRX8401 NA

Patients

Seq Age Sex Outcome Treatment
1 NA