ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-23743
- Event Type
- Malfunction
- Date Received
- August 19, 2014
- Report Date
- August 6, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION 10/09/2014-PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. REVIEW OF THE PUMP¿S BLACK BOX REVEALED UNEXPLAINED REBOOTING EVENTS. NO DAMAGE WAS OBSERVED TO THE BATTERY COMPARTMENT AND BATTERY CAP. THE BATTERY CAP CONTACT DIMENSIONS WERE WITHIN SPECIFICATION. THE BATTERY CAP WAS ABLE TO BE PROPERLY SECURED TO THE PUMP. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. NO EVIDENCE OF DAMAGE OR DEFECT WAS FOUND TO THE PUMP¿S INTERNAL COMPONENTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. IT WAS REPORTED THE PATIENT¿S BLOOD GLUCOSE WAS BETWEEN 250-499 MG/DL WITHOUT SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THE REPORTED BLOOD GLUCOSE EXCURSION DOES NOT MEET CRITERIA FOR A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 497336 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |