FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-05010
- Event Type
- Malfunction
- Date Received
- August 18, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 23, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND ONE GRIP WAS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .035 OFFSET AT THE TIPS. THE BENT GRIP DID NOT EXHIBIT SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT, HOWEVER LIKELY CAUSE OF BENDING REMAINS OVERLOADING AT THE TIP. FAILURE ANALYSIS CONCLUDED THAT THE BENT GRIP DAMAGE WAS LIKELY DUE TO MISHANDLING. FAILURE ANALYSIS ALSO OBSERVED THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .032 - .125 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE MAIN TUBE MATERIAL REMOVAL FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI GASTRIC SLEEVE RESECTION PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS PULLED OUT OF TRAY AND NOTICED THE GRASPERS WERE OFF CENTER. THE INSTRUMENT WAS TAKEN OFF FIELD AND ANOTHER WAS OPENED. NO FRAGMENTS FELL INTO THE PATIENT. THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT. NOTHING REPORTEDLY FELL INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 495362 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-07 | M10140318 026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |