FDA Adverse Event Malfunction Summary report: N

SOMATOM DEFINITION AS SYSTEM

MDR report key: 4013977 · Received May 16, 2014

Report

Report Number
2240869-2013-10287
Event Type
Malfunction
Date Received
May 16, 2014
Date of Event
November 3, 2013
Report Date
November 14, 2013
Manufacturer
SIEMENS AG MEDICAL SOLUTIONS
Product Code
JAK
PMA / PMN Number
K103127
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SIEMENS BECAME AWARE OF THE REPORTED EVENT ON (B)(6) 2013. THIS MDR IS BEING MAILED ON (B)(4) 2013. SIEMENS' PRELIMINARY INVESTIGATION REVEALED THAT THE BURNING ORIGINATED FROM THE CBR D11 (ELECTRONIC BOARD) AND THE POWER CABLES INSIDE THE IRSMX2C COMPUTER. THE COMPLETE IRSMC2C - FRU IRSMX2C COMPUTER HAS BEEN REPLACED. HOWEVER, SIEMENS' INVESTIGATION INTO THIS ISSUE IS ON-GOING AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

SIEMENS WAS NOTIFIED ON (B)(4) 2013, THAT ON (B)(6) 2013 AT APPROXIMATELY 23:00, THE CT SYSTEM SUDDENLY RESET ON ITS OWN. THEN, THE USER NOTICED A BURNING SMELL THAT WAS COMING FROM THE IRSMX2C AND THE SYSTEM WAS SHUTDOWN IMMEDIATELY. REPORTEDLY, THE BURNING SMELL SPREADS THROUGHOUT THE WHOLE GROUND FLOOR OF THE FACILITY AND EVERY WAS EVACUATED. THERE IS NO REPORT OF INJURY TO A PATIENT OR FACILITY PERSONNEL. THE REPORTED ISSUE OCCURRED IN MALAYSIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294635 SOMATOM DEFINITION AS SYSTEM SYSTEM, X-RAY, TEMOGRAPHY, COMPUTED JAK SIEMENS AG MEDICAL SOLUTIONS 08098027 NA

Patients

Seq Age Sex Outcome Treatment
1 SERVICE NO. (B)(4)| SW VERSION VA40A: TWO (2)| IRSMX2C COMPUTER: MAT NO. (1) 1026606 (2) 10162088