ECHELON LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-05752
- Event Type
- Malfunction
- Date Received
- August 15, 2014
- Report Date
- July 1, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RELEASE BUTTON. THE ANALYSIS FOUND THAT ONE PSE45A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR45W CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. FURTHERMORE, THE DEVICE WAS FOUND TO HAVE THE CLAMPING MECHANISM DAMAGED. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE CLAMP ARM RELEASE AND RELEASE BUTTON WERE NOTED WORN AND DAMAGED IN THE AREA WHERE BOTH INTERACT. THIS DAMAGE SUGGESTS THE DEVICE WAS FORCED TO OPEN WITH THE KNIFE NOT IN THE HOME POSITION BY PULLING THE CLOSING TRIGGER TO HOME. TO OPEN THE JAWS, SQUEEZE THE CLOSING TRIGGER, AND THEN SIMULTANEOUSLY PRESS THE ANVIL RELEASE SWITCH ON EITHER SIDE OF THE INSTRUMENT. WHILE PRESSURE IS STILL ON THE ANVIL RELEASE SWITCH, SLOWLY RELEASE THE CLOSING TRIGGER. NO FURTHER FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. HOWEVER, A DRY FIRE WAS PERFORMED AND THE FIRING MECHANISM WORKED AS INTENDED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT OPEN ALL THE WAY AND IT WOULD NOT LOCK. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO ADDITIONAL INFORMATION WAS AVAILABLE. THERE WAS NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 489871 | ECHELON LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L90G60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |