PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4013019
·
Received August 15, 2014
Report
- Report Number
- 1031452-2014-06244
- Event Type
- Malfunction
- Date Received
- August 15, 2014
- Report Date
- July 18, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER FAX, POWER SWITCH IS OPEN. PER INDEPENDENT REPAIR CENTER STATEMENT, UNIT WILL NOT START. THE KEY FAILURE WAS THE ON/OFF SWITCH FOR THE COMPONENTS PILOT VALVE WAS OPEN. ADDITIONAL MALFUNCTIONS WERE THE CASTER FOR THE BASE WAS MISSING, THE FITTING RECEPTACLE FOR THE BASE WAS MISSING AND THE SMALL CLAMP WAS LEAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 489718 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |