FDA Adverse Event Other Summary report: N

INFINIA HAWKEYE

MDR report key: 4011921 · Received December 2, 2013

Report

Report Number
9613299-2013-00069
Event Type
Other
Date Received
December 2, 2013
Date of Event
July 10, 2013
Report Date
July 10, 2013
Manufacturer
GE MEDICAL SYSTEMS F.I. HAIFA
Product Code
KPS
PMA / PMN Number
K022960
Removal / Correction Number
9613299-06/20/2013-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS ISSUE WAS IDENTIFIED AS PRESENTING A DIFFERENT FAILURE MODE THAN THE ONE IDENTIFIED IN MDR# 9613299-2013-00001. BASED ON ENGINEERING ANALYSIS A POTENTIAL GAP AS A RESULT OF PLASTIC SHIMS DOES NOT COMPROMISE THE STRUCTURAL INTEGRITY OF THE COMPONENT AND THEREFORE DOES NOT INTRODUCE A HAZARDOUS SITUATION. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

DURING SERVICE ACTIVITY RELATED TO A FIELD ACTION, A GE HEALTHCARE FIELD ENGINEER FOUND PLASTIC SHIMS UNDER THE RADIAL LINEAR RAIL. NO DETECTOR FALL EVENT AND NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
624180 INFINIA HAWKEYE SYSTEM, TOMOGRAPHY, COMPUTED, EMISSION KPS GE MEDICAL SYSTEMS F.I. HAIFA

Patients

Seq Age Sex Outcome Treatment
1