FDA Adverse Event Malfunction Summary report: N

INNOVATIVE PRODUCTS UNLIMITED, INC.

MDR report key: 4009408 · Received July 11, 2014

Report

Report Number
1833497-2014-00016
Event Type
Malfunction
Date Received
July 11, 2014
Date of Event
June 26, 2014
Report Date
July 10, 2014
Manufacturer
INNOVATIVE PRODUCTS UNLIMITED, INC.
Product Code
INM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE TWO CASTERS ON THE FRONT OF THE CHAIR WERE REPLACED WITH CASTERS THAT WERE NOT THE SAME AS THE ORIGINAL CASTERS. THE CASTERS IN THE FRONT WERE NOT THE SAME HEIGHT AS THE CASTERS IN THE BACK CAUSING STRESS ON THE PIPES BECAUSE THEY WERE NOT LINED UP CORRECTLY. THE FRONT TO BACK PIPE CAME LOOSE CAUSING THE T BELOW THE SEAT TO BREAK.

Description of Event or Problem · 1

PT WAS SITTING ON THE SHOWER CHAIR WHEN THE LEFT SIDE BROKE AT THE T FITTING LOCATED JUST UNDER THE SEAT. ALSO, THE LOWER SUPPORT PIPE GOING FROM THE REAR LEFT CASTER TO THE FRONT LEFT CASTER BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
406815 INNOVATIVE PRODUCTS UNLIMITED, INC. SHOWER CHAIR INM INNOVATIVE PRODUCTS UNLIMITED, INC. SC200 OS FS

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization