FDA Adverse Event Injury Summary report: N

SONICFILL COMPOSITE

MDR report key: 4006827 · Received August 13, 2014

Report

Report Number
2024312-2014-00577
Event Type
Injury
Date Received
August 13, 2014
Report Date
August 4, 2014
Manufacturer
KERR CORPORATION
Product Code
EBF
PMA / PMN Number
K091023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR'S OFFICE HAD INITIALLY ALLEGED THAT FOR AN UNKNOWN LOT NUMBER OF SONICFILL, APPROXIMATELY SEVEN (7) PATIENTS HAD EXPERIENCED SENSITIVITY AFTER PLACEMENT; HOWEVER, SPECIFIC PATIENT OR INCIDENT INFORMATION COULD NOT BE RECALLED. DURING A FOLLOW UP PHONE CALL ON 08/13/14, IT WAS DISCOVERED THAT TWO (2) ADDITIONAL LOT NUMBERS 5044576 AND 5091596 WERE ALSO ASSOCIATED WITH THE ALLEGED INCIDENT. ALTHOUGH THE OFFICE IDENTIFIED THREE (3) DIFFERENT LOTS ASSOCIATED WITH THE SENSITIVITY, THE OFFICE COULD NOT VERIFY WHICH LOT WAS USED ON ANY OF THE PATIENTS. A VISUAL EVALUATION WAS PERFORMED ON THE RETURNED PRODUCT FROM THE TWO (2) PROVIDED LOTS #5044576 AND #5091596, YIELDING RESULTS WITHIN SPECIFICATIONS. A DHR REVIEW INDICATED THAT THERE WERE NO DEVIATIONS FROM THE MANUFACTURING PROCESS. IN ADDITION, NO SIMILAR COMPLAINTS WERE RECEIVED WITH REGARD TO THESE LOTS.

Additional Manufacturer Narrative · 1

SPECIFIC PATIENT INFORMATION WITH REGARD TO GENDER, AGE, AND WEIGHT WAS NOT PROVIDED. THE PATIENT HAD EXPERIENCED SENSITIVITY AND THE DOCTOR REPLACED THE RESTORATIONS USING A DIFFERENT PRODUCT, WITHOUT FURTHER INCIDENT.. TO DATE, THE PATIENT HAS FULLY RECOVERED AND IS DOING FINE. THE PRODUCT INVOLVED IN THE ALLEGED INCIDENT WAS NOT RETURNED AND NO LOT NUMBER WAS PROVIDED; THEREFORE, NO EVALUATION CAN BE CONDUCTED.

Description of Event or Problem · 1

A DOCTOR'S OFFICE ALLEGED THAT APPROXIMATELY SEVEN (7) PATIENTS HAD EXPERIENCED SENSITIVITY AFTER PLACEMENT OF A RESTORATION USING THE SONICFILL PRODUCT. THIS IS THE FIFTH OF SEVEN (7) REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
481545 SONICFILL COMPOSITE TOOTH SHADE RESIN MATERIAL EBF KERR CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 Other| R PRIME & BOND