FDA Adverse Event Malfunction Summary report: N

ACTIVA

MDR report key: 4006517 · Received August 13, 2014

Report

Report Number
3004209178-2014-15256
Event Type
Malfunction
Date Received
August 13, 2014
Report Date
July 25, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
MHY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 37602, SERIAL# (B)(4), IMPLANTED: (B)(6) 2013, PRODUCT TYPE: IMPLANTABLE NEUROSTIMULATOR; PRODUCT ID 3389S-40, LOT# V109673, IMPLANTED: (B)(6) 2008, PRODUCT TYPE: LEAD; PRODUCT ID 7482A51, SERIAL# (B)(4), IMPLANTED: (B)(6) 2008, PRODUCT TYPE: EXTENSION; PRODUCT ID 3389S-40, LOT# V130326, IMPLANTED: (B)(6) 2008, PRODUCT TYPE: LEAD; PRODUCT ID 7482A51, SERIAL# (B)(4), IMPLANTED: (B)(6) 2008, PRODUCT TYPE: EXTENSION; PRODUCT ID 37642, SERIAL# (B)(4), PRODUCT TYPE: PROGRAMMER, PATIENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT¿S THERAPY WAS DOING WELL BUT IMPEDANCES HAD BECOME A CONCERN AND WORSENED IN (B)(6) 2013, POST IMPLANTABLE NEUROSTIMULATOR (INS) CHANGE-OUT. IMPEDANCE VALUES WERE AS LISTED: THERAPY IMPEDANCE WAS 2097 OHMS IN MARCH, THERAPY IMPEDANCE WAS ¿HIGH¿ IN (B)(6), ALL COMBINATIONS WITH 2 WERE ABOVE EXPECTED LEVELS IN (B)(6) 2014: C2 3996 OHMS, 02 5244 OHMS, 12 4638 OHMS, THERAPY IMPEDANCE WAS 1151 (NORMAL) IN (B)(6) 2014, AND THERAPY IMPEDANCE WAS ¿HIGH¿ ON (B)(6) 2014. THERE HAD BEEN NO REPORTS OF FALLS OR TRAUMA PRIOR TO CONCERN REGARDING IMPEDANCES. IT WAS NOTED THAT THE PATIENT DID STRETCHING, YOGA, AND DANCE. THE IMPEDANCE ISSUES SEEMED TO BE INTERMITTENT AND IT WAS STATED THAT IMPEDANCES WOULD BE CHECKED WHILE THE PATIENT WAS IN DIFFERENT POSITIONS THE NEXT TIME THE PATIENT WAS SEEN IN THE OFFICE. AT THE END OF THE CALL THERAPY WAS REPORTED TO BE DOING WELL AND THE IMPEDANCE PROBLEM WAS UNRESOLVED AT THAT TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
482200 ACTIVA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37602

Patients

Seq Age Sex Outcome Treatment
1 00061 YR