FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4002813 · Received August 12, 2014

Report

Report Number
2531779-2014-23176
Event Type
Malfunction
Date Received
August 12, 2014
Report Date
July 25, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS FOUND TO BE DIM, FADED AND DISCOLORED. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED BELOW THE FINGER PAD. ALSO UNRELATED TO THE COMPLAINT, THE DOWN ARROW BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONSIVE. THE KEYPAD COVER WAS FOUND TO BE INTACT. HOWEVER, WHEN THE KEYPAD COVER WAS REMOVED, CONTAMINATION WAS FOUND UNDER THE CONTRAST, UP AND DOWN BUTTON KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE DISTRIBUTER CONTACTED ANIMAS AND REPORTED A DIM, FADED AND COLOR SPECTRUM ISSUE WITH THE DISPLAY SCREEN. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED DURING TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
480248 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1