FDA Adverse Event
Malfunction
Summary report: N
2008K2 HEMODIALYSIS MACHINE
MDR report key: 4002443
·
Received June 10, 2014
Report
- Report Number
- 2937457-2014-01005
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- May 8, 2014
- Report Date
- May 12, 2014
- Manufacturer
- CONCORD MANUFACTURING
- Product Code
- KDI
- PMA / PMN Number
- K994267
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.
Description of Event or Problem · 1
A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THE AIR SEPARATOR AND THE O-RING WERE REPLACED. MACHINE IS BACK IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337551 | 2008K2 HEMODIALYSIS MACHINE | KDI | CONCORD MANUFACTURING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN SALINE (DISCARDED-NOT USED) |