FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4002441 · Received August 12, 2014

Report

Report Number
2032227-2014-07426
Event Type
Malfunction
Date Received
August 12, 2014
Date of Event
July 13, 2014
Report Date
July 13, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO UNEXPECTED SCREEN SWITCHING WAS NOTED. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES AND CRACKS ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP WAS GIVING HER TROUBLE WITH MOVING FROM ONE SCREEN TO THE NEXT. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THIS ISSUE. BLOOD GLUCOSE LEVEL WAS 130 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
477548 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAP A000290377

Patients

Seq Age Sex Outcome Treatment
1 57 YR