FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 4002402 · Received July 9, 2014

Report

Report Number
9615050-2014-04347
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
June 17, 2014
Report Date
June 18, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE ALARMED WITH A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE. FURTHER TESTING FOUND THAT THE DEVICE FAN WAS NOT ROTATING FAST ENOUGH DURING THE BATTERY CHARGING TO COOL THE DEVICE. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM WAS A BROKEN DEVICE COOLING FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM AN UNKNOWN FLOOR WITH A NOTE THAT INDICATED, TEMPERATURE TOO HIGH. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A S233 MALFUNCTION ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
399077 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK