LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-02098
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 3, 2014
- Report Date
- July 3, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 204) WAS CONFIRMED. AS RECEIVED, THE BELT FAILED INCOMING TESTING. UPON EVALUATION, THE YELLOW (PGND) WIRE WAS OPEN INSIDE THE TRUNK CABLE. THE CAUSE OF THE CODE 204 IS A DISRUPTION IN COMMUNICATION BETWEEN THE ELECTRODE BELT AND MONITOR. THE CAUSE OF THE DISRUPTION IN COMMUNICATION IS THE OPEN WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A SERVICE CODE 204. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392394 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |