FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 4002191
·
Received July 3, 2014
Report
- Report Number
- 3008642652-2014-02077
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 5, 2014
- Report Date
- July 2, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON INVESTIGATION, ALL OF THE CABLES OF THE ELECTRODE BELT WERE PULLED FROM THE STRAIN RELIEF, DAMAGING THE INTERNAL WIRES. THIS RESULTED IN AN INABILITY TO DETECT AND TREAT. THE ROOT CAUSE OF THE PULLED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SVC DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, ELECTRODE BELT SN (B)(4), THE BELT HAD A DAMAGED CABLE. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392283 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |