FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4002189 · Received August 12, 2014

Report

Report Number
2531779-2014-23169
Event Type
Malfunction
Date Received
August 12, 2014
Report Date
August 3, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 09/16/2014 WITH THE FOLLOWING FINDINGS: THE PUMP FAILED A LEAK TEST DUE TO A LEAK AT THE BATTERY COMPARTMENT, AND EVIDENCE OF MOISTURE INGRESS WAS OBSERVED INSIDE OF THE BATTERY COMPARTMENT; THE REPORTED EVENT WAS DUPLICATED. THE BATTERY COMPARTMENT WAS CRACKED, WHICH DIRECTLY CAUSED THE MOISTURE INGRESS EVENT. THE PUMP CASE WAS REMOVED, AND NO FURTHER EVIDENCE OF MOISTURE INGRESS WAS OBSERVED. A BATTERY CAP WAS NOT RETURNED WITH THE DEVICE; A TEST BATTERY CAP WAS USED DURING ALL STEPS OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. REPORTEDLY, EVIDENCE OF MOISTURE INGRESS WAS OBSERVED IN THE BATTERY COMPARTMENT; NO EVIDENCE OF DAMAGE TO THE PUMP WAS OBSERVED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
480154 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 46 YR