FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 4002004 · Received July 3, 2014

Report

Report Number
1824206-2014-01903
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 6, 2014
Report Date
June 6, 2014
Manufacturer
HILL-ROM INC.
Product Code
IOQ
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE LEFT SIDE RAIL LATCH WAS CONTAMINATED WITH A STICKY SUBSTANCE FROM DRIED FLUIDS THAT HAD PENETRATED THE SIDE RAILS CENTER ARM AND LATCHING MECHANISM ASSEMBLY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 THROUGH 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN CLEARED THE LEFT SIDE RAIL LATCHING MECHANISM TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THAT THE LEFT FOOT SIDE RAIL IS NOT LATCHING. THE BED WAS LOCATED ON THE 4TH FLOOR AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389060 TOTALCARE BARIATRIC BED AC POWERED ADJSUTABLE HOSPITAL BED IOQ HILL-ROM INC. 1840

Patients

Seq Age Sex Outcome Treatment
1