FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 4001043 · Received July 17, 2014

Report

Report Number
1828100-2014-00629
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PER THE USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED), DURING HIS TROUBLESHOOTING HE FOUND THAT THE FILTER/STRAINER WAS VERY DIRTY AND NOT ALLOWING ANY FLOW. THE BIOMED CLEANED THE FILTER/STRAINER AND THE COOLER HEATER UNIT WAS WORKING PROPERLY. NO PART TO BE RETURNED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THERE WAS LOW FLOW WITH THE COOLER HEATER UNIT. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
420753 TERUMO TCM II COOLING AND HEATING SYSTEM TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415