TERUMO TCM II COOLING AND HEATING SYSTEM
Report
- Report Number
- 1828100-2014-00629
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
PER THE USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED), DURING HIS TROUBLESHOOTING HE FOUND THAT THE FILTER/STRAINER WAS VERY DIRTY AND NOT ALLOWING ANY FLOW. THE BIOMED CLEANED THE FILTER/STRAINER AND THE COOLER HEATER UNIT WAS WORKING PROPERLY. NO PART TO BE RETURNED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THERE WAS LOW FLOW WITH THE COOLER HEATER UNIT. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 420753 | TERUMO TCM II COOLING AND HEATING SYSTEM | TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |