ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-23086
- Event Type
- Malfunction
- Date Received
- August 12, 2014
- Report Date
- August 1, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 10/15/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/01/2014 WITH THE FOLLOWING FINDINGS: THERE WAS NO EVIDENCE OF PUMP REBOOT IN THE BLACK BOX. THE BATTERY COMPARTMENT WAS CRACKED BETWEEN THE COVER AND BUMPER PAD. THERE WAS NO DAMAGE FOUND TO THE RETURNED BATTERY CAP AND IT'S MEASUREMENTS WERE WITHIN SPECIFICATIONS. THE RETURNED BATTERY CAP FULLY ATTACHED TO THE PUMP WITH NO YELLOW O RING SHOWING. NO POWER INTERRUPTIONS WERE DUPLICATED DURING A 24 HOUR DURATION TEST USING THE RETURNED BATTERY CAP. THERE WAS NO EVIDENCE OF INTERNAL MOISTURE OR DAMAGE TO THE POWER CIRCUIT. THE INITIAL COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. UNRELATED TO THE ALLEGATION OF A POWER INTERRUPTIONS, THE DISPLAY SCREEN WAS OBSERVED TO HAVE A PINKISH CONTRAST.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. REPORTEDLY, THE POWER ISSUE WAS OCCURRING DURING, OR IMMEDIATELY AFTER, A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 480419 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |