FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4000094 · Received August 11, 2014

Report

Report Number
2032227-2014-07350
Event Type
Malfunction
Date Received
August 11, 2014
Date of Event
July 12, 2014
Report Date
July 12, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACE. MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP. THE NUMBERS DID NOT RAMP UP ON ITS OWN OR SCROLL BAR MOVING WITH NO INPUT.

Description of Event or Problem · 1

CUSTOMER REPORTED THE UP ARROW BUTTON ON THE INSULIN PUMP BECAME UNRESPONSIVE. THE BLOOD GLUCOSE READING WAS 94 MG/DL. NOTHING FURTHER WAS REPORTED.