PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4000036
·
Received August 11, 2014
Report
- Report Number
- 3004209178-2014-92388
- Event Type
- Malfunction
- Date Received
- August 11, 2014
- Date of Event
- July 13, 2014
- Report Date
- July 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH SEVERELY SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND MISSING END CAP STICKER.
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
THE CALLER REPORTED THAT ONE OF THE BUTTONS ON THE INSULIN PUMP WAS NOT FUNCTIONING PROPERLY. THE CUSTOMER'S BLOOD GLUCOSE WAS 211 MG/DL. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.