FDA Adverse Event
Death
Summary report: N
MELKER EMERGENCY CUFFED SPECIAL OPERATIONS CRICOTHYROTOY CATHETER SET
MDR report key: 3999532
·
Received August 1, 2014
Report
- Report Number
- 1820334-2014-00350
- Event Type
- Death
- Date Received
- August 1, 2014
- Report Date
- July 16, 2014
- Manufacturer
- COOK INC
- Product Code
- BWC
- PMA / PMN Number
- K010016
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DATE OF DEATH UNK. DEVICE CODE - IT APPEARS THERE WAS NOTHING WRONG WITH THE SUPPLIED SCALPEL BUT THAT IT WAS NOT LONG ENOUGH TO SEE ON AN OBESE PT. NO MENTION OF PT SIZE IN THE IFU. EVENT EVALUATION: STILL UNDER INVESTIGATION.
Description of Event or Problem · 1
DURING A CRICOTHYROTOMY ON A LARGE (OVERWEIGHT) PT, THE INCLUDED DISPOSABLE SCALPEL WAS TOO SMALL TO CUT THROUGH ALL OF THE TISSUE IN ORDER TO CREATE AN AIRWAY. THE PT EXPIRED. THE USER FACILITY IS NOT CLAIMING THE DEVICE IS AT FAULT FOR THE PT OUTCOME. HOWEVER, IT WAS RECOMMENDED THAT A LARGER SCALPEL BE INCLUDED, TO ACCOMMODATE ALL TYPE OF PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449546 | MELKER EMERGENCY CUFFED SPECIAL OPERATIONS CRICOTHYROTOY CATHETER SET | BWC NEEDLE, EMERGENCY AIRWAY | BWC | COOK INC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |