HOMEPUMP ECLIPSE: 400 ML, 100 ML/HR
Report
- Report Number
- 2026095-2014-00118
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- June 12, 2014
- Manufacturer
- I-FLOW CORP.
- Product Code
- MEB
- PMA / PMN Number
- K052117
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, FR
- Reporter Occupation
- NURSE
Narratives
METHOD: AT THIS TIME ONLY ONE DEVICE HAS BEEN RECEIVED FOR AN ANALYSIS. WE ARE CURRENTLY PENDING THE RETURN OF THIS DEVICE. A REVIEW OF THE DEVICE HISTORY RECORDS (DHR) WAS PERFORMED FOR THE LOT NUMBER PROVIDED. RESULTS: AT THIS TIME THE EVALUATION IS STILL IN PROGRESS. RESULTS WILL BE PROVIDED ONCE COMPLETED. PER THE DHR REVIEW, THERE WERE NO REWORKS, SPECIAL CONDITIONS, OR RELATED NONCONFORMANCE REPORTS (NCRS) FOR THIS LOT. THE LOT MET THE PROCESS SPECIFICATIONS AT RELEASE. CONCLUSIONS: I-FLOW IS AWAITING THE RETURN OF THE TWO ADDITIONAL DEVICES THAT WERE REPORTED TO BE AVAILABLE FOR ANALYSIS. THE INVESTIGATION IS ONGOING, ONCE COMPLETED A FOLLOW UP REPORT WILL BE SUBMITTED. INFORMATION FROM THIS INCIDENT WILL BE INCLUDED IN OUR PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEM. ADDITIONAL INVESTIGATION MAY ARISE FROM ONGOING ANALYSIS, TREND INFORMATION, OR OTHER ANALYSIS AS APPROPRIATE.
FILL VOLUME: UNKNOWN. FLOW RATE: 100 ML/HR. PROCEDURE: UNKNOWN - ANP. CATHPLACE: UNKNOWN. PATIENT 2 OF 3. PLEASE REFERENCE 2026095-2014-00117/14-00495 (A), 2026095-2014-00119/14-00495 (C). IT WAS REPORTED THAT THREE PATIENTS EXPERIENCED FAST FLOWS WHILE USING A E401000 PUMP. THE NURSE DID NOT WAIT THE 4 HOURS BETWEEN FILLING THE PUMP AND INFUSION. PATIENT TWO'S PUMP INFUSED SOLU-MEDROL IN 2.5 HOURS INSTEAD OF THE EXPECTED 4 HOURS. ANP: ASKED NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 399222 | HOMEPUMP ECLIPSE: 400 ML, 100 ML/HR | ELASTOMERIC PUMP | MEB | I-FLOW CORP. | E401000 | 0200974832 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DRUG: SOLU-MEDROL LG DILUTED IN 400ML OF G5% |