FDA Adverse Event Malfunction Summary report: N

ENTREE CEILING MOUNT MICROSCOPE SYSTEM

MDR report key: 399756 · Received June 11, 2002

Report

Report Number
1937051-2002-00002
Event Type
Malfunction
Date Received
June 11, 2002
Date of Event
May 3, 2002
Report Date
June 2, 2002
Manufacturer
GLOBAL SURGICAL CORP.
Product Code
EPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IN 2002, THE CO RECEIVED A CALL FROM A CUSTOMER THAT THEIR CEILING MOUNTED MICROSCOPE FELL FROM THE CEILING. IT WAS INITIALLY REPORTED THAT NO ONE WAS INJURED. SUBSEQUENT REPORTS INDICATED THAT NO ONE WAS INJURED. SUBSEQUENT REPORTS INDICATED THAT IT HAD FALLEN ON A PT'S SHOULDER AND THAT PT HAD DEVELOPED A BRUISE. THE PT REFUSED ANY MEDICAL ATTENTION. AN ON-SITE TECHNICAL EVALUATION 4 DAYS LATER FOUND THAT THE COLUMN HAD BEEN IMPROPERLY INSTALLED IN ITS CLAMP. MECHANICAL MARKINGS INDICATED THAT COLUMN HAD BEEN ENGAGED INTO THE CLAMP ONLY ABOUT 1.5". THE DESIGN AND INSTALLATION INSTRUCTIONS INDICATE THAT THE COLUMN BE FULLY ENGAGED IN THE CLAMP AT 3". EXAMINATION OF A SECOND SYSTEM AT THE CUSTOMER'S SITE REVEALED THAT IT TO HAD BEEN IMPROPERLY INSTALLED IN THE SAME MANNER. BOTH SYSTEMS WERE THEN REINSTALLED PROPERLY INTO THE CLAMPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTREE CEILING MOUNT MICROSCOPE SYSTEM DENTAL MICROSCOPE EPT GLOBAL SURGICAL CORP. M 704CL NA

Patients

Seq Age Sex Outcome Treatment
1 NA