FDA Adverse Event
Malfunction
Summary report: N
LOANER CORE IMPACTION DRILL
MDR report key: 3994325
·
Received August 8, 2014
Report
- Report Number
- 0001811755-2014-02816
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- July 14, 2014
- Report Date
- July 14, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FAILURE FOR HEAT WAS CONFIRMED THROUGH FUNCTIONAL EVALUATION, DISASSEMBLY, AND VISUAL INSPECTION. THROUGH VISUAL INSPECTION, THE TECHNICIAN CONFIRMED THE BEARINGS WERE CORRODED. THE DEVICE WAS REPAIRED AND PLACED BACK INTO STRYKER STOCK.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE LOANER CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467787 | LOANER CORE IMPACTION DRILL | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |