FDA Adverse Event Malfunction Summary report: N

LOANER CORE IMPACTION DRILL

MDR report key: 3994325 · Received August 8, 2014

Report

Report Number
0001811755-2014-02816
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K112593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FAILURE FOR HEAT WAS CONFIRMED THROUGH FUNCTIONAL EVALUATION, DISASSEMBLY, AND VISUAL INSPECTION. THROUGH VISUAL INSPECTION, THE TECHNICIAN CONFIRMED THE BEARINGS WERE CORRODED. THE DEVICE WAS REPAIRED AND PLACED BACK INTO STRYKER STOCK.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE LOANER CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACK-UP EQUIPMENT WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467787 LOANER CORE IMPACTION DRILL DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO