FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3994287 · Received July 23, 2014

Report

Report Number
2518422-2014-01221
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE WAS CONTINUALLY ALARMING FOR A "CHECK CIRCUIT" CONDITION. THE DEVICE'S SYSTEM BOARD WILL BE REPLACED TO ADDRESS THE ISSUE. THE DEVICE WAS EVALUATED, BUT NOT YET REPAIRED.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431087 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054096

Patients

Seq Age Sex Outcome Treatment
1