FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3994266 · Received July 23, 2014

Report

Report Number
2518422-2014-01192
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 29, 2014
Report Date
June 29, 2014
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING TE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S POWER MANAGEMENT BOARD WAS REPLACED TO ADDRESS THE ISSUE. THE VENTILATOR FAILED STEPS DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUES.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431085 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054655