OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-20048
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Report Date
- July 31, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW-UP # 1 ¿ (10/08/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 9/5/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 9/29/2014 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(6), ALLEGING INACCURATE RESULTS OF 95 MG/DL ON THE SUBJECT METER AND 75 MG/DL ON A CALIBRATED LABORATORY DEVICE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S ACCURACY CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472494 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |