FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3994166 · Received August 8, 2014

Report

Report Number
2937094-2014-00704
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 13, 2014
Report Date
June 16, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/GLASS CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP AND THE METAL CAP ARE DETACHED AND RETURNED; THE GLASS FIBER EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE GLASS CAP EXHIBITS VERY MILD DETRITUS ADHESION. BASED ON THE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBERS METAL CAP CAME OFF. THE PROCEDURE WAS COMPLETED USING A SECOND SURGICAL FIBER. THERE WAS NO PATIENT INJURY REPORTED. GLAND VOLUME: 100 ML. TIME EXPENDED: 29:09. JOULES USED: 140,062.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467324 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 350B

Patients

Seq Age Sex Outcome Treatment
1