FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3994161
·
Received August 8, 2014
Report
- Report Number
- 2531779-2014-22839
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Report Date
- July 31, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/31/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE THREADS WERE DAMAGED. THE BATTERY CAP WAS NOT ABLE TO BE FULLY SECURED TO THE PUMP. THE BATTERY CAP THREADS WERE STRIPPED AND THE ¿COIN SLOT¿ WAS DAMAGED. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS WAS FOUND INSIDE THE PUMP. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT AND BATTERY CAP WERE DAMAGED. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 07/31/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467987 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |