FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3994128 · Received July 24, 2014

Report

Report Number
1828100-2014-00603
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SVC REP (FSR) REPLACED THE TWO FAILED OHMITE RESISTORS RESPONSIBLE FOR THESE SETTINGS. THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADD'L INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SVC REP (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE COOLER HEATER UNIT WOULD NOT OPERATE IN COOLING SPEED 1, 2 OR RECIRCULATE. THIS UNIT IS FOR ANIMAL RESEARCH, "NOT FOR HUMAN USE." SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433843 DUAL COOLER/HEATER SARNS DUAL COOLER/HEATER DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 11160

Patients

Seq Age Sex Outcome Treatment
1