FDA Adverse Event Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3994114 · Received August 8, 2014

Report

Report Number
2032227-2014-06616
Date Received
August 8, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

ALL OF THE BUTTONS FUNCTIONED PROPERLY. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. HOWEVER, CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT IS REPORTED THAT A CUSTOMER RECEIVED A BUTTON ERROR ALARM ON THEIR INSULIN PUMP. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS NOT RECORDED. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD LED TO THE BUTTON ERROR ALARM. THE CUSTOMER WAS INFORMED THAT THE PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. A NEW PUMP WAS SHIPPED TO THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.