FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3994068 · Received August 8, 2014

Report

Report Number
2649622-2014-08996
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 28, 2014
Report Date
June 28, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL ANALYSIS OF THE LEAD INDICATES DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO IMPLANT BUT NOT USED DUE TO THE TIP OF THE HELIX BEING BENT AND HELIX MECHANISM NOT FUNCTIONING CORRECTLY. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469606 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407658

Patients

Seq Age Sex Outcome Treatment
1 00088 YR