FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3994067 · Received August 8, 2014

Report

Report Number
2649622-2014-08994
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 4, 2014
Report Date
June 4, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 6947M62 LEAD IMPLANTED (B)(6) 2014. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED AND THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THERE WAS TISSUE ON THE HELIX.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD DISLODGED SEVERAL TIMES. THE LEAD WAS REMOVED. UPON EXPLANT, TISSUE WAS NOTED IN THE HELIX. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471404 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00051 YR Hospitalization| R DDBB1D4 ICD